5kW Flat Panel CArm Configuration Parameters |
1.Flat Panel Detector |
(1)Type of the detector:TFT monolithic amorphous siicon
(2)Panel size of the detector:210mm*210mm |
(3)Acquisition pixel matrix of the detector:1024×1024
(4)Pixel pitch of the detector:200 um |
(5)Spatial resolution of the detector:2.5LP/MM
(6)Pixel gray scale:16 Bit |
(7)Acquisition frame frequency:30fps |
2.High frequency and high voltage generator |
(1)Max.Output power:5kw |
(2)Input power supply:220 VAC |
(3)Photography voltage:40~125kV |
(4)Photography tube current range:10~100mA
(5)Exposure time range:1ms-5000ms
(6)mAs range:0.1mAs~200mAs
(7)Fluoroscopy voltage:40~120kV |
(8)Continuous fluoroscopy mA:0.5-5mA
(9)Pulsefluoroscopy mA:10-30mA |
(10)Pulse fluoroscopy frame frequency:1-30 frame
(11)Diagnostic self-test and display |
3.X-ray bulb tube assembly |
(1)Focus:0.3mm/0.6mm
(2)Max.KV:125kV |
(3)Anode target angle:12° |
4.C arm stand |
(1)Vertical displacement of C arm (electric):≥400mm
(2)Horizontal displacement of C arm (manual):≥200mm
(3)Rotation displacement of C arm (manual):≥±200° |
(4)Horizontal hunting displacement of C arm(manual):≥±12.5°
(5)Track displacement of C arm(manual):-30°±5°~+90°±5°
(6)Depth of arc of C arm≥650mm
(7)S.I.D:≥1000mm |
5.The gridand the beam limiting device |
(1)The gridmaterial:Aluminum based grid
(2)Size:23×23cm
(3)Grid ratio:100:1
(4)Wire/inch:215C |
(5)Grid Focallength:1000mm |
(6)Electric C arm DR beam limiting device |
6.Work station functions forimage acquisition,processing and diagnosis |
(1)Chinese user interface,standard DICOM3.0 image; |
10)Work station functions for imae acauisition adiust or breset window width/window level local automatic window level breset window width/window |
(3)Softwarepackage for the special image acquisition control and softwarepackage for special parts protocol processing; |
(4)Functions of natient manaement imaeacauisition imae brocesing limae correction imae fioing.USMsharnenine imaefiterinel.imae observation
15)Imae briting.DICoM orinting.baner nrinting.manual nrintine for the displaved imaces.sinele button for marking and nrinting the imaces.ontional for
(6)User personalization:showing the format and layout,default setting,toolbar setting,parts protocol enhancement filter;
(7)Image display:display configuration 1920×1080,HD display. |
8.Work station |
(1)CPU:Inteli7,3GHz or above; |
(2)Host memory:≥16GB DDR31600 high speed memory; |
(3)Hard disk:1T/7200rpm large capacity and high speed hard disk;
(4)Work station monitor:IPS LCD monitor; |
(5)Network card and network interface:1000M network card,1000M network interface,Jumbo Frames:9K;
(6)DICOM3.0 interface; |
(7)Windows 1064 bit |